CUT
Cut I
A general cutting with a standard RF waveform used in open and endoscopic surgical procedures, Cut I is for use in a wide range of tissue with minimal hemostasis.
Cut II
Tailored specifically for Laparoscopic procedures, Cut II offers a precisely managed cutting energy that limits output for minimizing the potential risk of capacitive coupling and alternate site burns due to weakened insulation.
Blend (Level 1, 2, 3 & 4)
Blended cutting offers varying levels of hemostasis during a cut procedure providing increased control of bleeding at the surgical application site.
- Level 1: 75% Duty Cycle
- Level 2: 62.5% Duty Cycle
- Level 3: 50% Duty Cycle
- Level 4: 37.5% Duty Cycle
COAGULATION
Pinpoint
A general coagulation mode delivering rapid desiccation effects with minimal tissue damage.
Spray
Recommended for hemostasis in highly vascular or large surface areas and for non-contact tissue coagulation to control superficial bleeding. Spray mode permits dual, simultaneous activation of two monopolar accessories with audible dual activation tone alert.
Gentle
Gentle mode offers more concentrated coagulation than in other modes. When coagulation is necessary in short amounts of time, reduced electrode carbonization is provided.
Bovie’s Cut II is tailored specifically for Laparoscopic procedures. The mode offers a precisely managed cutting energy that limits output thereby minimizing the potential risk of capacitive coupling and alternate site burns.
The SurgiCenter|PRO offers simultaneous activation in spray with audible activation tones. The simultaneous delivery produces equal and even power outputs in each accessory, for a safer delivery of dual monopolar energy.
The power can be adjusted up or down in variable steps while performing a procedure, to increase procedural efficiency and provide the appropriate amount of power to the surgical site.
The SurgiCenter | PRO recalls the last used power and mode. The surgeon can store 10 user-defined RF presets for easy recall.